AI Troubleshooting

Diagnose GC-MS chromatography and ionization problems

M

Symptom-Based Diagnosis

Pick what you're observing on the instrument

Likely Causes

  • Active sites (exposed silanols/metal) in the inlet liner or column reacting with polar analytes
  • Column contamination from non-volatile matrix residue
  • Analyte thermally degrading or adsorbing at the inlet before it reaches the column
  • Injection technique — needle dwell time too long, or a cold-spot in the liner

Recommended Fixes

  • Replace the inlet liner (deactivated, single-taper for splitless) — liners are consumables and the #1 source of tailing
  • Install a guard/retention-gap column ahead of the analytical column
  • Lower the inlet temperature toward the analyte's boiling point, or switch to PTV/cold injection for thermally labile compounds
  • Derivatize free -OH/-COOH/-NH groups (silylation) to cap active sites causing the interaction

Or ask the AI Scientist directly

Hi — I'm your GC-MS AI Scientist. Ask me about method development, EI/CI source parameters, SRM optimization, or troubleshooting.