Symptom-Based Diagnosis
Pick what you're observing on the instrument
Likely Causes
- Active sites (exposed silanols/metal) in the inlet liner or column reacting with polar analytes
- Column contamination from non-volatile matrix residue
- Analyte thermally degrading or adsorbing at the inlet before it reaches the column
- Injection technique — needle dwell time too long, or a cold-spot in the liner
Recommended Fixes
- Replace the inlet liner (deactivated, single-taper for splitless) — liners are consumables and the #1 source of tailing
- Install a guard/retention-gap column ahead of the analytical column
- Lower the inlet temperature toward the analyte's boiling point, or switch to PTV/cold injection for thermally labile compounds
- Derivatize free -OH/-COOH/-NH groups (silylation) to cap active sites causing the interaction
Or ask the AI Scientist directly
Hi — I'm your GC-MS AI Scientist. Ask me about method development, EI/CI source parameters, SRM optimization, or troubleshooting.